The Avery Berkel L126 is a robust retail scale and indicator system used widely for trade-legal weighing. To maintain its accuracy and compliance with trade standards, regular calibration is essential. This guide covers the procedures for calibrating and resetting the L126, based on technical service standards. Before starting, ensure you have the following ready:
Remove the weight and press the Test button to restart the unit. The scale will go through its warm-up routine and return to zero. How to Perform a Zero Reset avery berkel l126 calibration
Press to scroll to "Zero Balance," then press Semi-Auto Tare to capture the current zero without clearing existing span data. Troubleshooting and Maintenance YouTube·Velocity How to restart Avery Berkel scale The Avery Berkel L126 is a robust retail
Locate the SW1 button (often a small micro-switch) usually found on the bottom of the unit or inside the casing on the left side of the main board. Press it using a small tool like a pen. The display should change to "DIAG" (Diagnose). Before starting, ensure you have the following ready:
Press the Zero button twice until the display shows "CAL" (Calibration). Confirm your selection by pressing the Test or Store button.
The display should read "DONE" or "PASS" once the span is captured.