Fmx002 Manual Top !!better!! — Simco

If the display shows , the charge exceeds 20kV. Immediately move the meter away to avoid damaging the sensitive sensor. 5. Maintenance and Troubleshooting

The FMX-002 features two red focusing lights. Move the meter closer or further until the two red spots merge into a single point on the surface. This ensures you are at the calibrated distance of 1 inch (25mm) . Read the Display: Blue Bar: Indicates negative polarity. Red Bar: Indicates positive polarity. Digital Readout: Shows the kilovolts (kV) present.

Move the meter slowly toward the target object. simco fmx002 manual top

If you are looking for the "manual top" essentials—the critical operating procedures and maintenance tips—this guide covers everything you need to get accurate readings. 1. Overview of the Simco FMX-002

The FMX-002 typically operates in two modes depending on the charge level: Measures up to ±1.49kV with high resolution. If the display shows , the charge exceeds 20kV

Automatically switches to measure up to ±20.0kV when higher charges are detected.

If you are not grounded, the meter will measure your body's potential rather than the target's. Zeroing the Meter Turn the power on. Maintenance and Troubleshooting The FMX-002 features two red

Clip the other end to a verified earth ground (like a grounded equipment frame or workbench).