Diagnostica Stago Start 4 Coagulation Analyzer Manual May 2026

Place the cuvette with the sample and the metal ball into the incubation station. Start: Move the cuvette to the measurement channel.

If you are looking for specific , calibration curves , or reagent-specific volumes for your STart 4,

To ensure the longevity of your STart 4, follow these routine maintenance steps outlined in the manual: diagnostica stago start 4 coagulation analyzer manual

Verify the temperature of the incubation blocks using a certified thermometer to ensure it remains at 37°C (+/- 0.5°C). ⚠️ Troubleshooting Common Issues Error Message / Problem Potential Cause No Clot Detected Low fibrinogen or missing ball Check cuvette for metal ball; re-run sample. Temperature Fluctuation Room draft or faulty sensor Move away from AC vents; contact technical support. Pipette Sync Error Loose connection Check the RS232 or specialized pipette cable connection. 📋 Technical Specifications Measurement Channels: 4 independent channels.

Wipe down the incubation area with a mild disinfectant. Ensure no reagent spills have occurred in the measurement channels. Weekly: Check the pipette tip and tubing for clogs. Place the cuvette with the sample and the

Turn the switch at the rear. The system will perform a self-test.

The heart of the STart 4 is the . Unlike optical systems that can be affected by lipemic, icteric, or hemolyzed samples, the STart 4 uses a mechanical approach: ⚠️ Troubleshooting Common Issues Error Message / Problem

Each cuvette requires one metal ball for the VBDS system to function.